Standard stock
Cstock = mass×potency×dry fraction / volume
Use consistent mass and volume units and document the basis of the assigned potency.

The 30-day usage period has ended.
Prepare standards, calculate every level, fit the line, inspect residuals, and determine unknown concentration.
Use consistent mass and volume units and document the basis of the assigned potency.
The same formula gives the aliquot required for a target concentration.
x is concentration, y is response, m is slope, and b is intercept.
High r or R² alone does not prove acceptable linearity.
Apply the full sample dilution factor after back-calculation.
Residuals should be inspected for patterns, curvature, or changing variance.
This app reports back-calculated concentration deviation for each non-zero level.
Use cautiously when the predicted response is close to zero.
| Level | Nominal x | Observed y | Predicted ŷ | Residual y−ŷ | Residual % | Back-calculated x | Level deviation % |
|---|