As-found calibration result
Direct definition
Measurement evidence of an instrument's condition before a calibration adjustment or repair at the service event.
Detailed technical explanation
As-found data document pre-intervention indication errors at selected calibration points and enable retrospective exposure assessment. They must be retained even if an adjustment restores compliance.
Practical example
ILLUSTRATIVE — At a hypothetical GMP site, a controlled pressure transmitter was replaced during a production shutdown. The site evaluates objective evidence, intended use and product risk rather than treating a calibration sticker as definitive proof.
Why it matters
They are central to evaluating potentially affected batches after an unexpected drift excursion.
Pharmaceutical applications
They are central to evaluating potentially affected batches after an unexpected drift excursion.
Common misconception
A passing as-left result makes the earlier as-found failure irrelevant.
Source references — not yet verified
- Source register — 21 CFR 211.160(b)(4) — Laboratory controls and calibration
Source supplied with the draft; version and applicability not verified.
Source register link — unverified applicability - Source register — 21 CFR 211.68(a) — Automatic, mechanical, and electronic equipment
Source supplied with the draft; version and applicability not verified.
Source register link — unverified applicability
Review and applicability
Scientific review incomplete. No scientific reviewer, approval date or regulatory endorsement is claimed.
Educational information. Apply the relevant jurisdiction, product context and approved site quality system.