Pharmaceutical Calibration, Metrology and Measurement System Control · Definition

Measurement precision

Direct definition

Closeness of agreement between independent measurement indications or measured values obtained under specified conditions.

Detailed technical explanation

Precision can be evaluated through repeatability, intermediate precision or reproducibility, each defined by personnel, environment, time and instrument conditions. It does not establish freedom from systematic error.

Practical example

ILLUSTRATIVE — At a hypothetical GMP site, a pH electrode shows growing bias across three documented calibration cycles. The site evaluates objective evidence, intended use and product risk rather than treating a calibration sticker as definitive proof.

Why it matters

Precision evidence supports capability assessment for balances, pipettes and process sensors.

Pharmaceutical applications

Precision evidence supports capability assessment for balances, pipettes and process sensors.

Common misconception

Good precision guarantees that an instrument has no bias.

Source references — not yet verified

  • Source register — 21 CFR 211.160(b)(4) — Laboratory controls and calibration

    Source supplied with the draft; version and applicability not verified.

    Source register link — unverified applicability
  • Source register — 21 CFR 211.68(a) — Automatic, mechanical, and electronic equipment

    Source supplied with the draft; version and applicability not verified.

    Source register link — unverified applicability
  • Source register — JCGM 200:2012 — International Vocabulary of Metrology (VIM), 3rd ed.

    Source supplied with the draft; version and applicability not verified.

    Source URL requires editorial verification.

  • Source register — NIST — Measurement Uncertainty

    Source supplied with the draft; version and applicability not verified.

    Source URL requires editorial verification.

Review and applicability

Scientific review incomplete. No scientific reviewer, approval date or regulatory endorsement is claimed.

Educational information. Apply the relevant jurisdiction, product context and approved site quality system.

Related definitions and questions

Measurement accuracyMeasurement biasWhat is the difference between accuracy and precision?Why does balance calibration matter for dispensing?Can an instrument be accurate but not precise?How should a balance minimum-weight capability be evaluated?How should a laboratory evaluate pipette performance across volumes?A dispensing balance passes at 200 g but fails near a 2 g working load. Can it remain in use for 2 g API dispensing?A release assay uses a pipette that later fails a low-volume check. How should the impact assessment be structured?

More on this topic

Metrological traceabilityCalibrationMeasurement uncertaintyTolerance and maximum permissible errorDecision rule and guard bandCalibration interval

Related resource collections

GuidelinesSOP LibraryResourcesAsk Pharma AI — separate approved-source collection